摘 要:本文通過火焰原子吸收法對土壤中速效鉀的含量進行測量。分析測量過程中各不確定因素的來源,并對其進行量化。計算、合成標準不確定度和擴展不確定度。最終得出結論:土壤中速效鉀的含量為200.28mg/kg。影響鉀含量的主要不確定因素主要是標準曲線擬合的不確定度。最終實驗結果表示為:(200.28±5.6)mg/kg。 關鍵詞: 火焰原子吸收,鉀,土壤,不確定度
Abstract:This paper focuses on the measurement of rapid available potassium content in soil by FAAS. The sources of uncertainty in the process of measurement were analyzed and calculated and the combined uncertainty and expanded uncertainty were obtained. Then the final conclusion is drawn that the measurement of rapid available potassium content in soil is 200.28mg/kg. The main uncertainty came from and the standard curve fitting combined uncertainty is. The final experimental can be expressed as (200.28±5.6)mg/kg. Key words: FAAS; potassium; soil; uncertainty
國外對于測量不確定度的研究與應用要早于我國,GUM在各國校準和測試實驗室得到了廣泛應用。范圍也極其廣泛,工業、商業部門,政府部門,科研院所等都有大量的研究和應用。在化學測量不確定度方面,目前發表的報道多是標準物質的測量不確定度的評定,制備人血清、南極洲標準物質、Li同位素混合物、硝酸同位素溶液 (Jean.ClaudeWblff,1998)、硅酸鹽玻璃標準物質(R.w.Hinton,1997)等。最早在中國從事測量不確定度研究的學者是中國計量科學研究院研究員,同時也是國際測量不確定度工作組成員的劉智敏先生。實例中以基本標準物質研究以及物理測試為主。在我國,不確定度理論也應用到各個測量領域的具體實際工作中,如計量、環境保護、醫療衛生、食品檢驗、冶金、儀器儀表、系統工程和電子技術等部門都有相當的研究人員在做相關的研究工。 |